The GX-L/GF-L series, the successor to A&D’s long-selling GP series, continues to allow precise weighing of heavy or large amounts of materials in demanding conditions. In addition, now as part of the A&D Apollo family, these models also offer various advanced features that make your weighing easier and more efficient, not to mention more accurate.
Capacity | 102 kg |
---|---|
Readability | 1 g |
Repeatability (Standard Deviation) | 1 g |
Minimum Weight *ii |
1.4 kg |
Linearity | ±2 g |
Stabilization Time | Approximately 1.5 second (typical when set to FAST) |
Sensitivity Drift | ±6 ppm / °C (10°C to 30°C / 50°F to 86°F, when automatic self sensitivity adjustment is OFF) |
Accuracy Immediately After Internal Sensitivity Adjustment *iii |
±10 g |
Operating Environment | 5°C to 40°C / 41°F to 104°F, 85% RH or less (no condensation) |
Data Memory | 200 weighing results + 50 sensitivity adjustment/calibration test results (with timestamps), 50 unit masses for counting mode, 20 sets of upper/lower limit values for comparator mode, and 20 tare values |
Display Refresh Rate | 5 times/second, 10 times/second or 20 times/second |
Units of Measure *iv |
g (gram), kg (kilogram), oz (ounce), lb (pound), lb-oz (pound-ounce), ozt (troy ounce), ct (metric carat), mom (momme), dwt (pennyweight), pcs (counting mode), % (percent mode), DS (density mode), and a user-programmable unit |
Counting Mode | Minimum unit mass: 1 g Number of samples: 10, 25, 50 or 100 pieces |
Percent Mode | Minimum 100% reference mass: 100 g % readability: 0.01%, 0.1%, 1% (depends on the reference mass stored) |
Communication Interface | RS-232C (D-Sub 9-pin) and USB (mini B) |
Applicable Calibration Weight Value | 10 kg, 20 kg, 40 kg, 50 kg, 60 kg, 80 kg x 100 kg |
Weighing Pan Size | 386 mm x 346 mm (15.2" x 13.6") |
External Dimensions (W x D x H) *v |
Swing-arm type: 373 mm x 615 mm x 130 mm (14.68" x 24.21" x 5.12") |
Net Weight | Approximately 18 kg (39.7 lb) |
Dust and Waterproof Rating | IP65 |
Power Supply / Consumption | AC adapter / approximately 30 VA |
*i Smart range function: Automatically switches between the precision and standard ranges, and recovers the full precision range with a press of the TARE key.
*ii Pursuant to the United States Pharmacopeia (USP), Chapter 41
*iii Under a stable environment: No rapid temperature/humidity change, vibration, draft, magnetism, static, etc., between 10 °C and 30 °C (50 °F and 86 °F). The mass of the internal weight may vary with age.
*iv One additional unit from tael (Singapore/HK jewelry/Taiwan), tola, or Newton can be added upon request.
*v Dimensions when the swing-arm stand is pulled fully down.
*vi Dimensions of the base unit. The dimensions of the display unit is 260 (W) × 164 (D) × 202 (H) mm, and the cable length is 3 m.