Secura gives you the security and peace of mind of knowing that you have done everything right. Besides providing highly accurate weighing results and operating convenience, Secura also features built-in protection systems for complete reliability and regulatory compliance - the safe and secure way. Real-time guidance prompts for leveling, automatic internal adjustment and 100% traceable, clear documentation with sample and batch identifiers make your lab work more efficient.
Capacity | 610 g |
---|---|
Readability | 1 mg |
Repeatability (standard deviation) | 1 mg |
Linearity | 2 mg |
Typical Starting point of the Operating Range * | 1.5 g |
Optimal Starting Point of the Operating Range * | 0.82 g |
Sensitivity Drift between +10°C and +30° C | ± 2 ppm/K |
Typical Stabilization Time | 1 seconds |
Calibration | isoCAL internal calibration |
isoCAL Settings | Temperature Change: 2 K Time Interval: 6 h |
Display Result (depending on the filter level) | 0.1 / 0.2 s |
Weighing Pan Size | ø 120 mm (ø 4.75") |
Weighing Chamber Height ** | 209 mm |
Leveling | Intelligent, optoelectronic leveling sensor with alarm function and interactive user guidance for reliable leveling |
Selectable Weight Units | Gram, kilogram, carat, pound, ounce, troy ounce, Hong Kong tael, Singapore tael, Taiwan tael, grain, pennyweights, milligram, parts per pound, China tael, mommes, Austrian carat, tola, baht, mesghal and Newton (limited for verified models) |
Interface |
Mini USB
|
Display | Touch screen with Sartorius graphical user interface |
Standard Built-in Applications | Weighing, Density, Percentage, Checkweighing, Peak Hold, Counting, Unstable Conditions (Animal Weighing) |
APC Features |
|
Anti-theft Lock | Kensington lock and lockdown capability for cable or chain (not included) |
Net Weight (approximately) | 5.1 kg |
Dimensions (D x W x H) | 359 mm x 214 mm x 319 mm (14.1" x 8.4" x 12.6") |
* According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity. Depending on the installation location and environmental conditions, the value could be higher.
** Upper edge of the weighing pan to the lower edge of the upper draft shield panel